Table 2 Field performance of the SD BIOLINE HIV/Syphilis Duo dual rapid test in Haiti for detection of Treponema pallidum antibodies using a dual HIV/syphilis test, 2014

From: Dual rapid lateral flow immunoassay fingerstick wholeblood testing for syphilis and HIV infections is acceptable and accurate, Port-au-Prince, Haiti

 

Treponema pallidum reference test*

Total

Sensitivity (95 % CI)

Specificity (95 % CI)

Kappa Coefficient (95 % CI)

Pos

Neg

    

SD BIOLINE HIV/Syphilis Duo test

Pos

109

17

126

96.5 % (91.2 %, 99.0 %)

90.8 % (85.7 %, 94.6 %)

.85

Neg

4

168

172

  

(.79, .91)

Total

113

185

298

   
  1. *Reference test was conducted using Treponema Pallidum Hemaglutination Assay (TPHA) (Human Gesellschaft fur Biochemica und Diagnostica mbH, Wiesbaden, Germany). TPHA indeterminate results were retested using a Treponema pallidum enzyme-linked immunosorbent assay test (ELISA)(Architect Syphilis TP; Abbott, Wiesbaden, Germany) and were reclassified based on the ELISA result