Skip to main content

Table 1 Analytical sensitivity and precision comparisons of the manual and automated methods.

From: Automated extraction protocol for quantification of SARS-Coronavirus RNA in serum: an evaluation study

 

Sensitivity comparison

Precision comparison

SARS-CoV RNA Concentration (copies/mL)

10

102

103

103

105

    

mean

S.D.

CV (%)

mean

S.D.

CV (%)

Automated

1/10a

7/10

10/10

349

268

76.9

35052

17892

51

Manual

4/10

9/10

10/10

1539

697

45.3

83591

39606

47.4